Istruzioni per l'uso A&D, Modello MS-70

Produttore : A&D
File Size : 1.2 mb
File Nome :
Lingua di insegnamento: en
Vai a scaricare



Facilità d'uso


PROG 2 0. 000g 6.3.2. Recalling a Measurement Program with a Program Number Stored measurement programs can be recalled with a program number. PROG "1 0 . 000g 1. Press and hold the PROGRAM key in the gram display. 2. Press the v or ^ key to select a program number PROG "2Program number 3. Press the ENTER key to use the selected number. MS-70 / MX-50 / MF-50 / ML-50 25 end PROG "2 0.000g 7. Measurement Programs 7.1. List of Measurement Programs 7. Measurement Programs 7.1. List of Measurement Programs There are five analysis modes. Measurement Programs Analyzing mode to Finish Measurement Drying Program Heating pattern Drying Temperature MS-70, MX-50, MF-50 ML-50 MS-70 MX-50 MF-50 ML-50 Standard Mode 5td Measurement condition is automatically selected by ACCURACY and minimum value of % display. When drying rate is less than preset termination value, measurement is completed automatically. Standard drying Ramp drying Step drying Standard drying 30°C to 200°C 50°C to 200°C Quick Mode quc Quick drying Automatic Mode U-a When drying rate is less than preset termination value, measurement is completed automatically. Standard drying Ramp drying Step drying Standard dryingTimer Mode U-t Sample is dried for a preset time. 1min. to 480min. Manual Mode U-m Measurement is completed by the STOP key. Max. 480 min. Drying rate: Change of moisture content per one minute [%/min] 26 MS-70 / MX-50 / MF-50 / ML-50 7.1.1. ACCURACY of the Standard Mode and Quick Mode ACCURACY of measurement can be set either HI , MID. or LO . The sample quantity is automatically selected by ACCURACY. The termination value of the analyzing mode is automatically selected by ACCURACY and minimum scale value of % display. The analyzing mode is the program to finish the measurement when a change of moisture content per one minute becomes smaller than a preset termination value. The settings are as follows: Specify an ACCURACY. Drying rate: Change of moisture content per one minute [%/min] Standard Mode ACCURACY Model Minimum scale LO 0.001 % 0.01 %/min 0.02 %/min 0.05 %/min 0.01 % 0.02 %/min 0.05 %/min 0.10 %/min MS-70 0.1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.01 % 0.02 %/min 0.05 %/min 0.10 %/min MX-50 0.1 % 0.10 %/min 0.20 %/min 0.05 % 0.05 %/min 0.10 %/min 0.1 % MF-50 1 % 0.1 % Preset Termination value ML-50 1 % 0.10 %/min 0.20 %/min 0.50 %/min Sample quantity 10 g 5 g 1 g Use Precise result - Quick measurement Quick Mode ACCURACY Model Minimum scale LO 0.001 % 0.02 %/min 0.05 %/min 0.05 %/min 0.01 % 0.05 %/min 0.10 %/min 0.20 %/min MS-70 0.1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.01 % 0.05 %/min 0.10 %/min 0.20 %/min MX-50 0.1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.05 % 0.1 % MF-50 1 % 0.10 %/min 0.20 %/min 0.50 %/min 0.1 % Preset Termination value ML-50 1 % 0.20 %/min 0.50 %/min 1.00 %/min Sample quantity 5 g 2 g 1 g Use Precise result - Quick measurement MS-70 / MX-50 / MF-50 / ML-50 27 7.1.2. Analyzing mode of the Automatic Mode When the change of moisture content per one minute is less than the preset value, the measurement is automatically completed. RangePreset Termination value to complete measurement MS-70 MX-50 MF-50 ML-50 2.00 %/min 1.00 %/min 0.50 %/min Available Available 0.20 %/min Available range range 0.10 %/min Available range (Factory setting) (Factory setting) 0.05 %/min range (Factory setting) 0.02 %/min (Factory setting) 0.01 %/min 0.005 %/min 0.002 %/min 0.001 %/min Unavailable item 7.1.3. Analyzing mode of the Timer Mode Sample is dried for a preset analysis time. Drying Time 1 minute to 480 minutes. Setting interval : 1 minute during 1 minute to 60 minutes. 5 minutes during 60 minutes to 480 minutes. Factory setting: 10 minutes. 7.1.4. Drying Program (Heating Pattern and Drying Temperature) Heating pattern Standard Mode, Automatic Mode, Manual Mode Timer Mode Quick Mode Standard drying Temperature Temperature Time - Ramp drying Temperature Time Temperature Time1 Time2 - Step drying Stage 1 Temperature1 Time Temperature2 Stage 2 Temperature1 Time1 Temperature2 Stage 1 Stage 2 Time2 - Quick drying -- Temperature 200°C 3 min. "Temperature 1" can be set higher than "temperature 2" in step drying. ML-50, can select standard drying and quick drying only, does not display heating pattern. 28 MS-70 / MX-50 / MF-50 / ML-50 How to Select quick mode (Extracts) Press the SELECT quc key during the gram display. Display using the v or ^ key. ML-50 is does not display heating pattern. Quick mode "q"u"c How to Select a Heating pattern without quick mode (Extracts) Press the SELECT key to select temperature during the gram display. "1"0"5°C Press the PROGRAM key to select a symbol of heating pattern are displayed in order. ML-50 can select standard drying and quick drying only. Heating pattern Display examples during settings Temperature Temperature Time Heating pattern Drying Temperature at the Sample Pan Drying temperature range for MS-70.......................................30°C to 200°C, 1°C interval. Drying temperature range for MX-50, MF...

Questo manuale è adatto anche per i modelli :
Termostato - MF-50 (1.2 mb)
Termostato - ML-50 (1.2 mb)
Termostato - MF-50 (1.2 mb)
Termostato - MX-50 (1.2 mb)

Scrivi la tua opinione del dispositivo



Il messaggio
Il tuo nome :
Inserire le due cifre :
capcha





categorie