Log:
valutazioni - 4, GPA: 4.5 ( )

Istruzioni per l'uso Bacharach, Modello Fyrite Tech 60

Produttore : Bacharach
File Size : 400.58 kb
File Nome :
Lingua di insegnamento: en
Vai a scaricare



Facilità d'uso


Allow the probe to cool before handling (about 5 minues). 10. End a test by first removing the probe from the flue-gas stream, and then allowing the pump to run until the O2 reading reaches approximately 20.9%. Turning OFF the Analyzer 11. Turn OFF the analyzer by pressing the I/O button. There is a 5-second delay before the analyzer actually turns OFF, during which time the unit can be turned back ON by pressing the ENTER button. A Fyrite Tech 60 will not turn OFF if a high level of CO is still being detected. The pump will remain running and the message “PurG CO” is displayed until the detected CO level drops below 50 ppm. This purge process can be bypassed by again pressing the I/O button. Screen Displays - Fyrite Tech 50 Use the ▲ and ▼ buttons to scroll through the screens. Screen Displays - Fyrite Tech 60 Use the ▲ and ▼ buttons to scroll through the screens. Fuel Select: F1 = Natural Gas F2 = Oil #2 F3 = LPG F4 = Kerosene O, Ambient Air Temperature Stack Temperature & Efficiency Excess Air Fuel Select: F1 = Natural Gas F2 = Oil #2 F3 = LPG F4 = Kerosene CO & O, CO Air Free Ambient Air Temperature Stack Temperature & Efficiency Excess Air CO, & CO, & - 2 - - 3 - - 4 -...

Questo manuale è adatto anche per i modelli :
Depuratori d - Fyrite 24-9429 (400.58 kb)
Depuratori d - Fyrite Tech 50 (400.58 kb)

Scrivi la tua opinione del dispositivo



Il messaggio
Il tuo nome :
Inserire le due cifre :
capcha





categorie